Using Turbidity Secondary Standards

Secondary standards give customers a fast, safe way to check that their turbidity meters are operating correctly, without the cost or complexity of using a primary standard. In some cases, they can also be used to recalibrate a turbidity meter.

TurbSense secondary standards are produced with approximate turbidity values, known as nominal values, of 0.9 NTU and 18 NTU. The actual turbidity value of each standard may vary slightly from these nominal values due to small differences in materials and manufacturing. For this reason, every secondary standard is measured at the point of manufacture, with the measured turbidity value clearly marked on the base of the standard and inside the case lid.

Run time: 2.19 min

How to Use Turbidity Secondary Standards from Pi - Measurements - TurbSense

Other Videos by this Supplier

Tell Us What You Think

Do you have a review, update or anything you would like to add to this video content?

Leave your feedback
Your comment type
Submit

Sensors Videos by Subject Matter

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.